Highlights:
|
Orphines are a class of synthetic opioids originally developed in Belgium in the 1960s for their potent anaesthetic properties and later withdrawn from clinical use due to their overdose risk. It is now increasingly found in the illicit drug supply, with some analogues having comparable to substantially more potency levels than fentanyl. Brorphine was the first member reported in the illicit drug market in 2019, and additional analogues, including cychlorphine and spirochlorphine, have emerged more recently. Following China's class-wide scheduling of nitazene compounds in July 2025, orphine analogues have proliferated in the illicit drug supply a pattern consistent with the earlier shift from fentanyl to nitazenes.
In the United States and Canada, orphines have been detected in samples sold as other opioids, such as hydromorphone and oxycodone, that contained no other active drug. As with other potent synthetic opioids, even small amounts of orphines can cause respiratory depression, and unknowingly consuming orphine-containing substances increases the risk of accidental overdose. Early evidence suggests that existing fentanyl and nitazene test strips do not reliably detect orphines, creating a gap in point-of-use drug checking that this test strip is designed to address.
The Rapid Response® Orphine Test Strip from BTNX is specifically designed for the swift screening of Orphines and Orphine analogues in liquid/powder samples. The test incorporates an antibody that selectively identifies orphine and orphine analogues from other substances. Like other harm reduction test strips, it operates on a competitive binding principle and generates a colorimetric result. If any of the listed orphine substances are present in the sample at a concentration surpassing the designated cut-off, a singular-colored line will appear in the test.
Positive - Orphine Detected
Only one coloured line appears in the control region (C). No apparent coloured line appears in the test region (T).
Negative - Orphine Could Not Be Detected
Two coloured lines appear on the membrane. One line appears in the control region (C) and another line appears in the test region (T).
Analytical Sensitivity
The following compounds were individually spiked into water to evaluate the detection capability of the Rapid Response® Orphine Test Strip. The cut-off listed in the table represents the concentration at which a substance will begin to show a mix of positive and negative results.
| Compounds | Cut Off |
| N-Propionitrile Chlorphine | 2000 ng/mL |
Analytical Specificity
The following compounds were spiked into water at the concentrations shown to examine possible cross-reactivity with the Rapid Response® Orphine Test Strip. None produced a false-positive result.
| Compounds | Concentration | Result |
| Xylazine | 10 mg/mL | Negative |
| Diphenhydramine | 10 mg/mL | Negative |
| Caffeine | 5 mg/mL | Negative |
| Lidocaine | 10 mg/mL | Negative |
| Phenacetin | 0.1 mg/mL | Negative |
| Levamisole | 10 mg/mL | Negative |
| Acetaminophen | 10 mg/mL | Negative |
| Dimethylsulphone | 10 mg/mL | Negative |
The Rapid Response® Orphine Test Strip (Liquid / Powder) is available to order under the product code ORP-18S26-100 from Lochness Medical Supplies in the USA, and BTNX Inc. in Canada and the international market.
Canada